INTRODUCTION — Portable monitoring (PM) is a diagnostic test for obstructive sleep apnea (OSA), a disorder characterized by repetitive episodes of apnea or reduced inspiratory airflow due to upper airway obstruction during sleep. It has evolved as an alternative to overnight, attended, in-laboratory polysomnography (PSG) in selected patients.
Advantages of PM include its convenience (it can be performed in the patient's home or in a hospital room) and its lower costs (PM devices are less costly than complete polysomnography systems and the attendance of a technologist is not required). However, a disadvantage is that fewer physiologic variables are measured during PM than PSG, which can lead to misinterpretation of the results. Other advantages and disadvantages are listed in the figure (table 1).
The United States Centers for Medicare and Medicaid Services (CMS) have released guidelines that state that results from PM can be used to support a prescription for positive airway pressure therapy [1]. The American Academy of Sleep Medicine (AASM) has also released clinical practice guidelines to guide clinicians in the use of PM [2,3].
The use of PM in the diagnostic evaluation of suspected OSA is reviewed here. The diagnostic approach to a patient with suspected OSA is described separately. (See "Clinical presentation and diagnosis of obstructive sleep apnea in adults".)
TYPES OF MONITORING DEVICES — Four types of sleep study monitoring devices have been defined [2,4].
Type 1 devices — Type 1 monitoring devices are used for in-laboratory, technician-attended, overnight polysomnography (PSG). These devices are not used for PM and are discussed separately. (See "Polysomnography in obstructive sleep apnea in adults".)
Type 2 devices — Type 2 monitoring devices can record the same variables as type 1 devices. The major difference from type 1 devices is that they can be used outside of the sleep laboratory and a technologist is not present during the recording.
Type 3 devices — Type 3 monitoring devices typically measure four physiologic variables, including two respiratory variables (eg, respiratory movement and airflow), a cardiac variable (eg, heart rate or an electrocardiogram), and arterial oxyhemoglobin saturation via pulse oximetry. Some devices have additional signals that can detect snoring, determine body position, or detect movement. Sleep variables (eg, sleep stages, sleep continuity) are typically not measured by a type 3 device, although some newer devices have methods for estimating sleep. A technologist is not present during the recording.
Recordings from a typical type 3 portable monitoring device are shown in the figure (figure 1). In this example, respiratory effort is detected by a chest wall impedance monitor, airflow is detected by a heat-sensitive thermistor at the nose and mouth, and oxyhemoglobin saturation is measured by a pulse oximeter. The heart rate is also recorded. Measurement of these variables is generally sufficient to detect most apneas (obstructive, central, mixed) and hypopneas.
Type 3 devices have important limitations:
Type 4 devices — Type 4 monitoring devices are defined differently by different organizations. The American Academy of Sleep Medicine (AASM) defines type 4 monitoring devices as devices that record one or two variables (eg, arterial oxyhemoglobin saturation and airflow) and can be used without a technician. These devices are called continuous single or dual bioparameter devices. In contrast, the United States Centers for Medicare and Medicaid Services (CMS) guidelines include devices that measure three variables as type 4 monitoring devices. A technologist is not present during the recording.
Type 4 devices that record one or two variables provide limited information. Pulse oximetry and airflow are the physiological variables that are most commonly measured. As a result, derived information typically includes the frequency of apneas, frequency of hypopneas, AHI, baseline oxyhemoglobin saturation (SpO2), mean SpO2, frequency of oxyhemoglobin desaturation, duration of oxyhemoglobin desaturation, degree of oxyhemoglobin desaturation, and nadirs of SpO2. All of the limitations of type 3 devices described above also apply to type 4 devices.
A novel type 4 device measures arterial tonometry, oximetry, snoring, and body position. It uses an automated algorithm to detect breathing events based upon peripheral arterial tone [5,6].
Pulse oximetry alone is also considered a type 4 device, which is discussed further below. (See 'Pulse oximetry' below.)
INDICATIONS — The most common indications for PM include the diagnostic evaluation of suspected OSA and titration of positive airway pressure therapy.
Diagnosis — Many PM devices have been validated against standard PSG, typically by testing the same patient with both modalities in the sleep laboratory [7-10]. A wide range of diagnostic characteristics have been reported using PM, but the sensitivity and specificity seem to be high in populations considered by sleep specialists to be at high risk of OSA on the basis of clinical symptoms, assuming there are no comorbid medical disorders or sleep disorders [2].
We endorse the most recent clinical practice guidelines from the American Academy of Sleep Medicine (AASM) [2,3]:
The CMS guidelines do not make any specific recommendations about which type of PM device should be used, nor do they specify the appropriate population to undergo PM. However, they do indicate that positive airway pressure therapy prescriptions will be covered by Medicare and Medicaid only if OSA is diagnosed using a type 1, 2, or 3 device [1]. In addition, a Type 4 device that measures at least three variables is acceptable.
Limitations — PM has important limitations that should be routinely considered by the clinician who is interpreting the results of the study:
Titration — The role of portable monitoring in determining the initial level of positive airway pressure therapy is described in detail elsewhere. (See "Initiation of positive airway pressure therapy for obstructive sleep apnea in adults", section on 'Determining the amount of positive airway pressure'.)
Reassessment — PM can be used to assess the adequacy of the prescribed therapy within a patient's usual environment and to guide adjustment of the therapy. The approach is similar to that for diagnosis and titration, respectively.
PULSE OXIMETRY — Pulse oximetry is a widely accepted and important component of both polysomnography (PSG) and PM. However, when it is measured alone or with only one other variable, it is not recommended for the diagnostic evaluation of suspected OSA [1,2,17]. In this section, we briefly describe the reason that overnight oximetry alone is not considered an acceptable diagnostic modality.
The characteristics of overnight pulse oximetry alone as a diagnostic test are highly dependent upon whether the criteria used to define a positive test are quantitative or qualitative:
Overnight pulse oximetry alone can be either a sensitive or a specific test for OSA, but not both. As a result, either false-positive or false-negative tests will be common, depending on the criteria chosen to define a positive test.
SUMMARY AND RECOMMENDATIONS
All topics are updated as new information becomes available. Our peer review process typically takes one to six weeks depending on the issue.