Official reprint from UpToDate® www.uptodate.com
©2012 UpToDate®
The content on the UpToDate website is not intended nor recommended as a substitute for medical advice, diagnosis, or treatment. Always seek the advice of your own physician or other qualified health care professional regarding any medical questions or conditions. The use of this website is governed by the UpToDate Terms of Use (click here) ©2012 UpToDate, Inc.
Portable monitoring in obstructive sleep apnea in adults
All topics are updated as new evidence becomes available and our peer review process is complete.
Literature review current through: Apr 2012. | This topic last updated: Aug 11, 2011.

INTRODUCTION — Portable monitoring (PM) is a diagnostic test for obstructive sleep apnea (OSA), a disorder characterized by repetitive episodes of apnea or reduced inspiratory airflow due to upper airway obstruction during sleep. It has evolved as an alternative to overnight, attended, in-laboratory polysomnography (PSG) in selected patients.

Advantages of PM include its convenience (it can be performed in the patient's home or in a hospital room) and its lower costs (PM devices are less costly than complete polysomnography systems and the attendance of a technologist is not required). However, a disadvantage is that fewer physiologic variables are measured during PM than PSG, which can lead to misinterpretation of the results. Other advantages and disadvantages are listed in the figure (table 1).

The United States Centers for Medicare and Medicaid Services (CMS) have released guidelines that state that results from PM can be used to support a prescription for positive airway pressure therapy [1]. The American Academy of Sleep Medicine (AASM) has also released clinical practice guidelines to guide clinicians in the use of PM [2,3].

The use of PM in the diagnostic evaluation of suspected OSA is reviewed here. The diagnostic approach to a patient with suspected OSA is described separately. (See "Clinical presentation and diagnosis of obstructive sleep apnea in adults".)

TYPES OF MONITORING DEVICES — Four types of sleep study monitoring devices have been defined [2,4].

Type 1 devices — Type 1 monitoring devices are used for in-laboratory, technician-attended, overnight polysomnography (PSG). These devices are not used for PM and are discussed separately. (See "Polysomnography in obstructive sleep apnea in adults".)

Type 2 devices — Type 2 monitoring devices can record the same variables as type 1 devices. The major difference from type 1 devices is that they can be used outside of the sleep laboratory and a technologist is not present during the recording.

Type 3 devices — Type 3 monitoring devices typically measure four physiologic variables, including two respiratory variables (eg, respiratory movement and airflow), a cardiac variable (eg, heart rate or an electrocardiogram), and arterial oxyhemoglobin saturation via pulse oximetry. Some devices have additional signals that can detect snoring, determine body position, or detect movement. Sleep variables (eg, sleep stages, sleep continuity) are typically not measured by a type 3 device, although some newer devices have methods for estimating sleep. A technologist is not present during the recording.

Recordings from a typical type 3 portable monitoring device are shown in the figure (figure 1). In this example, respiratory effort is detected by a chest wall impedance monitor, airflow is detected by a heat-sensitive thermistor at the nose and mouth, and oxyhemoglobin saturation is measured by a pulse oximeter. The heart rate is also recorded. Measurement of these variables is generally sufficient to detect most apneas (obstructive, central, mixed) and hypopneas.

Type 3 devices have important limitations:

Type 4 devices — Type 4 monitoring devices are defined differently by different organizations. The American Academy of Sleep Medicine (AASM) defines type 4 monitoring devices as devices that record one or two variables (eg, arterial oxyhemoglobin saturation and airflow) and can be used without a technician. These devices are called continuous single or dual bioparameter devices. In contrast, the United States Centers for Medicare and Medicaid Services (CMS) guidelines include devices that measure three variables as type 4 monitoring devices. A technologist is not present during the recording.

Type 4 devices that record one or two variables provide limited information. Pulse oximetry and airflow are the physiological variables that are most commonly measured. As a result, derived information typically includes the frequency of apneas, frequency of hypopneas, AHI, baseline oxyhemoglobin saturation (SpO2), mean SpO2, frequency of oxyhemoglobin desaturation, duration of oxyhemoglobin desaturation, degree of oxyhemoglobin desaturation, and nadirs of SpO2. All of the limitations of type 3 devices described above also apply to type 4 devices.

A novel type 4 device measures arterial tonometry, oximetry, snoring, and body position. It uses an automated algorithm to detect breathing events based upon peripheral arterial tone [5,6].

Pulse oximetry alone is also considered a type 4 device, which is discussed further below. (See 'Pulse oximetry' below.)

INDICATIONS — The most common indications for PM include the diagnostic evaluation of suspected OSA and titration of positive airway pressure therapy.

Diagnosis — Many PM devices have been validated against standard PSG, typically by testing the same patient with both modalities in the sleep laboratory [7-10]. A wide range of diagnostic characteristics have been reported using PM, but the sensitivity and specificity seem to be high in populations considered by sleep specialists to be at high risk of OSA on the basis of clinical symptoms, assuming there are no comorbid medical disorders or sleep disorders [2].

We endorse the most recent clinical practice guidelines from the American Academy of Sleep Medicine (AASM) [2,3]:

  • PM for diagnostic evaluation of suspected OSA should be performed only in conjunction with a comprehensive sleep evaluation, preferably by a sleep medicine specialist.
  • PM may be used as an alternative to PSG for the diagnosis of OSA in patients with a high pre-test probability of moderate to severe OSA. (See "Clinical presentation and diagnosis of obstructive sleep apnea in adults".)
  • PM should not be used in patients who have comorbid medical conditions that predispose to sleep related breathing disorders or in whom another sleep disorder is suspected. As an example, PM should not be used in patients with heart failure because they are predisposed to Cheyne-Stokes breathing. (See "Cheyne-Stokes breathing and obstructive sleep apnea in heart failure".)
  • Ideally, PM must record airflow, respiratory effort, and blood oxygenation. Alternatively, devices which measure peripheral arterial tone, oximetry, and snoring are also considered adequate.
  • An experienced sleep technician, sleep technologist, or appropriately trained provider must apply the PM sensors or directly educate the patient about the correct application of sensors. In addition, there should be methodology in place to monitor the quality of recordings.
  • PM devices must be capable of displaying the raw data for review by the clinician, in order to allow assessment of the quality of the data.
  • All patients undergoing PM should receive a follow-up visit with an appropriately trained provider to discuss the results of the test.

The CMS guidelines do not make any specific recommendations about which type of PM device should be used, nor do they specify the appropriate population to undergo PM. However, they do indicate that positive airway pressure therapy prescriptions will be covered by Medicare and Medicaid only if OSA is diagnosed using a type 1, 2, or 3 device [1]. In addition, a Type 4 device that measures at least three variables is acceptable.

Limitations — PM has important limitations that should be routinely considered by the clinician who is interpreting the results of the study:

  • Type 3 and type 4 devices often underestimate the AHI (described above), increasing the likelihood of a false-negative result. As a result, the pretest clinical suspicion for OSA and the results of PM studies should always be considered together. Additional testing is indicated whenever there is a reasonable clinical suspicion for OSA, but the PM study is negative.
  • Type 4 devices that record only one or two physiologic variables cannot distinguish obstructive and central events, in part because respiratory effort is not measured. Patients at risk for central apneas, hypopneas, or hypoventilation should not be tested with such devices.
  • Different manufacturers use different sensors and potentially different algorithms to identify respiratory events. Many devices do not allow the raw data to be viewed by the provider who ordered the test because the scoring algorithm is proprietary. This is problematic according to numerous studies that showed that manual scoring or manual editing of automated scoring improves diagnostic accuracy compared to automated scoring alone [11-16]. Clinical practice guidelines recommend manual scoring or manual editing of automated scoring.

Titration — The role of portable monitoring in determining the initial level of positive airway pressure therapy is described in detail elsewhere. (See "Initiation of positive airway pressure therapy for obstructive sleep apnea in adults", section on 'Determining the amount of positive airway pressure'.)

Reassessment — PM can be used to assess the adequacy of the prescribed therapy within a patient's usual environment and to guide adjustment of the therapy. The approach is similar to that for diagnosis and titration, respectively.

PULSE OXIMETRY — Pulse oximetry is a widely accepted and important component of both polysomnography (PSG) and PM. However, when it is measured alone or with only one other variable, it is not recommended for the diagnostic evaluation of suspected OSA [1,2,17]. In this section, we briefly describe the reason that overnight oximetry alone is not considered an acceptable diagnostic modality.

The characteristics of overnight pulse oximetry alone as a diagnostic test are highly dependent upon whether the criteria used to define a positive test are quantitative or qualitative:

  • Overnight pulse oximetry tends to have a high specificity but a low sensitivity when quantitative criteria are used [18-20]. As an example, a study of 200 patients with suspected OSA compared overnight pulse oximetry versus polysomnography [19]. Positive overnight pulse oximetry was defined as a decrease in the oxyhemoglobin saturation (SpO2) of 4 percent or greater from baseline, to a value 90 percent or lower. Overnight pulse oximetry had a sensitivity of 41 percent and a specificity of 97 percent.
  • In contrast, overnight pulse oximetry tends to have a high sensitivity but a low specificity when qualitative criteria are used [9]. Qualitative criteria emphasize pattern recognition. The overnight pulse oximetry strip is printed and visually analyzed for a pattern of multiple, short-duration fluctuations of the SpO2. No specific level of desaturation is required for a positive study. As an example, a study of 240 patients with suspected OSA compared overnight pulse oximetry versus polysomnography. Positive overnight pulse oximetry was defined as greater than 10 fluctuations of SpO2 per hour of sleep. The sensitivity and specificity of overnight pulse oximetry were 98 and 48 percent, respectively.

Overnight pulse oximetry alone can be either a sensitive or a specific test for OSA, but not both. As a result, either false-positive or false-negative tests will be common, depending on the criteria chosen to define a positive test.

SUMMARY AND RECOMMENDATIONS

  • Portable monitoring (PM) is an acceptable approach to the diagnostic evaluation of suspected obstructive sleep apnea (OSA). However, it should only be used for the diagnostic evaluation of suspected OSA in patients with a high pre-test probability of having moderate to severe OSA and no comorbid medical or sleep disorders. (See 'Diagnosis' above.)
  • PM can also be used to guide the titration of positive airway pressure therapy if the chosen mode of positive airway pressure is either continuous positive airway pressure (CPAP) or autotitrating positive airway pressure. (See 'Titration' above.)
  • Four types of PM devices have been defined on the basis of the amount of data that they can measure. Type 3 and 4 devices have highly variable diagnostic performance and most do not include a conventional measure of sleep, which has many drawbacks. (See 'Types of monitoring devices' above.)
  • Pulse oximetry is a widely accepted and important component of both polysomnography and PM. However, it should NOT be used alone for the diagnostic evaluation of suspected OSA. (See 'Pulse oximetry' above.)

Use of UpToDate is subject to the Subscription and License Agreement.

REFERENCES

  1. Department of Health and Human Services, Center for Medicare and Medicaid Services. Decision Memo for Continuous Positive Airway Pressure (CPAP) Therapy for Obstructive Sleep Apnea (OSA). CAG#0093R. March 13, 2008.
  2. Collop NA, Anderson WM, Boehlecke B, et al. Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 2007; 3:737.
  3. Epstein LJ, Kristo D, Strollo PJ Jr, et al. Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults. J Clin Sleep Med 2009; 5:263.
  4. Ferber R, Millman R, Coppola M, et al. Portable recording in the assessment of obstructive sleep apnea. ASDA standards of practice. Sleep 1994; 17:378.
  5. Choi JH, Kim EJ, Kim YS, et al. Validation study of portable device for the diagnosis of obstructive sleep apnea according to the new AASM scoring criteria: Watch-PAT 100. Acta Otolaryngol 2010; 130:838.
  6. Pang KP, Gourin CG, Terris DJ. A comparison of polysomnography and the WatchPAT in the diagnosis of obstructive sleep apnea. Otolaryngol Head Neck Surg 2007; 137:665.
  7. Man GC, Kang BV. Validation of a portable sleep apnea monitoring device. Chest 1995; 108:388.
  8. White DP, Gibb TJ, Wall JM, Westbrook PR. Assessment of accuracy and analysis time of a novel device to monitor sleep and breathing in the home. Sleep 1995; 18:115.
  9. Sériès F, Marc I, Cormier Y, La Forge J. Utility of nocturnal home oximetry for case finding in patients with suspected sleep apnea hypopnea syndrome. Ann Intern Med 1993; 119:449.
  10. Tonelli de Oliveira AC, Martinez D, Vasconcelos LF, et al. Diagnosis of obstructive sleep apnea syndrome and its outcomes with home portable monitoring. Chest 2009; 135:330.
  11. Ancoli-Israel S, Mason W, Coy TV, et al. Evaluation of sleep disordered breathing with unattended recording: the Nightwatch System. J Med Eng Technol 1997; 21:10.
  12. Cirignotta F, Mondini S, Gerardi R, et al. Unreliability of automatic scoring of MESAM 4 in assessing patients with complicated obstructive sleep apnea syndrome. Chest 2001; 119:1387.
  13. Dingli K, Coleman EL, Vennelle M, et al. Evaluation of a portable device for diagnosing the sleep apnoea/hypopnoea syndrome. Eur Respir J 2003; 21:253.
  14. Øverland B, Bruskeland G, Akre H, Skatvedt O. Evaluation of a portable recording device (Reggie) with actimeter and nasopharyngeal/esophagus catheter incorporated. Respiration 2005; 72:600.
  15. Yin M, Miyazaki S, Ishikawa K. Evaluation of type 3 portable monitoring in unattended home setting for suspected sleep apnea: factors that may affect its accuracy. Otolaryngol Head Neck Surg 2006; 134:204.
  16. Yin M, Miyazaki S, Itasaka Y, et al. A preliminary study on application of portable monitoring for diagnosis of obstructive sleep apnea. Auris Nasus Larynx 2005; 32:151.
  17. Chesson AL Jr, Berry RB, Pack A, et al. Practice parameters for the use of portable monitoring devices in the investigation of suspected obstructive sleep apnea in adults. Sleep 2003; 26:907.
  18. Williams AJ, Yu G, Santiago S, Stein M. Screening for sleep apnea using pulse oximetry and a clinical score. Chest 1991; 100:631.
  19. Douglas NJ, Thomas S, Jan MA. Clinical value of polysomnography. Lancet 1992; 339:347.
  20. Gyulay S, Olson LG, Hensley MJ, et al. A comparison of clinical assessment and home oximetry in the diagnosis of obstructive sleep apnea. Am Rev Respir Dis 1993; 147:50.
Topic 7694 Version 4.0

All topics are updated as new information becomes available. Our peer review process typically takes one to six weeks depending on the issue.